28
so the meter will be cleaned and disinfect about 11000 cycles
that include cleanings plus disinfections (1 cycle = 1 cleaning
+ 1 disinfection) during its 5 years life cycle.
1. Before the test, the user should clean and wash hands. If you
are a caregiver, (not the patient), the please use disposable
gloves.
2.
After test, take one wipe to clean the meter’s housing, buttons,
lens, mark, usb cover and support block, including its front and
back for 10 seconds for cleaning until there is no soil on the
surface. Then disinfect the meter’s housing, buttons, lens, mark,
usb cover and support block use another wipe for 2 min, keep
the surface wet for 2 min.
3. Take off the glove; discard carefully according to your health-
care provider’s instruction.
Note:
1. Each disinfection step requires a pre-cleaning step.
2. If the meter is being operated by a second person who is pro-
viding testing assistance to the user, the meter and lancing
device shall be cleaned and disinfected first.
3.
The users should wash hands thoroughly with soap and water
after handling the meter, lancing device, or test strips.
4. If you have any questions you can dial the number
1-877-342-2385
5. Please refer to the following linkage for detail information:
“FDA Public Health Notification: Use of Fingerstick Devices
on More than One Person Poses Risk for Transmitting
Bloodborne Pathogens: Initial Communication” (2010)
http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/
ucm224025.htm
“CDC Clinical Reminder: Use of Fingerstick Devices on More
than One Person Poses Risk for Transmitting Bloodborne
Pathogens” (2010) http://www.cdc.gov/injectionsafety/Fingerstick
-DevicesBGM.html
6. Table:
A list of the signs of potential deterioration
(Physical and in performance)
What do you see?
What to do?
The buttons do Stop using and contact local customer
not work
services or place of purchase for
assistance, or toll free service hotline:
1-877-342-2385
The device does Stop using and contact local customer
not work services or place of purchase for assi-
stance, or toll free service hotline:
1-877-342-2385
Discoloration of the Stop using and contact local customer
buttons , meter services or place of purchase for assi-
casing or lancing stance, or toll free service hotline:
device 1-877-342-2385
Corrosion, crazing, Stop using and contact local customer
embrittlement, crac- services or place of purchase for assi-
king of the buttons, stance, or toll free service hotline:
meter casing, display 1-877-342-2385
or lancing device
Cloudiness of the Stop using and contact local customer
lens services or place of purchase for assi-
stance, or toll free service hotline:
1-877-342-2385
1. Model: CLEVER CHOICE HD (with “Clever Choice” test strip
and Clever Choice control solution)
2. Machine size: 88mm (L)×55mm (W)×26mm(H)
3. Weight: 45g (excluding batteries)
4. Measuring method: Amperometric technology using glucose
oxidase